In-office biological indicator (BI) system certified for use with all steam sterilization cycles operating at 121-135°C.
- For use with all steam sterilizers
- Fast results with spore growth observable in as little as 2 1/2 hrs. with 24 hr. final incubation
- Media growth promotion validation per USP
- Allows facilities to comply with industry guidelines for sterilizer monitoring
Each SporView Self-Contained Steam BI (SCBI) contains Geobacillus stearothermophilus spores and growth media along with a chemical indicator label printed with lot and expiration date. After processing, the SCBI is activated and then incubated at 55-60°C for 24 hours.
Special Note: Media growth promotion has been validated for growth of G. stearothermophilus organism after exposure to a steam cycle of 132°C for 20 minutes exposure.
ANSI/AAMI standards recommend steam sterilizers be biologically monitored at least weekly, preferably daily and every load with an implant. CSA standards recommend BI testing daily and every load with an implant. CDC guidelines recommend BI testing weekly and every load with an implant.
Heat Sensitive | No |
CE Marked | No |
Manufacturer Item Number | SCS-025 |
Country of Origin | United States |
Hazardous Shipping Restrictions | Not Hazardous |
Perishable | No |
Refrigeration Required | No |
Latex Free | Yes |
UOM | 25/BOX |